Stage
|
Cutoff
|
AUC (95% CI)
|
Sensitivity (%)
|
Specificity (%)
|
PPV (%)
|
NPV (%)
|
P value*
|
|---|
|
F ≥ 2
|
> 11.3
|
0.950 (0.803–0.996)
|
94.1
|
84.6
|
84.2
|
90.9
|
< 0.001
|
|
F ≥ 3
|
> 18.1
|
0.924 (0.766–0.989)
|
83.3
|
100.0
|
100.0
|
96.0
|
< 0.001
|
- 2D-SWE = two-dimensional shear-wave elastography. Diagnostic accuracy of each variable in association with fibrosis stage. The performance of the selected best cutoff values was indicated. AUC = the area under the receiver operating curve. PPV, positive predictive value; NPV, Negative predictive value; *Determined using receiver operating characteristic curve analysis